background image

4.1 Practical notes

818 IC Pump / 8.818.1023 Instructions for Use

23

4

Notes – Maintenance – Faults

4.1 Practical

notes

4.1.1

Protection against foreign particles

To protect the column against foreign particles which could have an

adverse influence on the separation efficiency, we advise you to subject

both the eluents and all samples to microfiltration (0.45

µ

m filter) and

to siphon the eluent through the 6.2821.090 Aspirating filter.
To avoid contamination by abrasive particles arising from piston seals

of the 818 IC Pump, it is advantageous to install an in-line filter 24

between the pump and the 820 IC Separation Center (see Fig. 5). We

recommend to use the 6.2821.100 Filter unit PEEK which is included

in the standard accessories of the 818 IC Pump (see section 2.3.3).

4.1.2 Pulsation

dampener

To protect the column material against pressure shocks caused by

injection we recommend to use a pulsation dampener between the 818

IC Pump and the 820 IC Separation Center (see Fig. 5). The optional

6.2620.150 Pulsation dampener MF (see section 2.3.2) is eminently

suitable for this purpose.

4.1.3 Eluents

Treatment

For the preparation of the eluents one should use chemicals of a purity

degree of at least "p.a.". For dilution please use only high purity water.
Fresh eluents should always be microfiltered (0.45

µ

m filter) and

degassed (with N

2

, He or vacuum). For in-line degassing of eluents the

837 IC Eluent Degasser is available from Metrohm. The eluent should

be continuously stirred with a magnetic stirrer, particularly when the

recycling procedure is employed or when alkaline eluents are used. For

alkaline eluents and eluents with low buffering capacity one should

preferably use CO

2

absorbers.

The supply vessel containing the eluent must be closed as tightly as

possible to avoid excessive evaporation. This is primarily important with

eluents containing organic solvents (e.g. acetone), the evaporation of

which can lead to drifts in the long term. If work is performed in a very

sensitive range, even if one drop of condensate falls back in the eluent

this can cause a noticeable change in the background conductivity.

Precipitates

Salt crystals between the piston and the seal are the cause of abrasive

particles, which can enter the eluent. These lead to contaminated