Metrohm 818 IC Pump User Manual

Page 27

Advertising
background image

3.3 Settings in the "818 IC Pump" window

818 IC Pump / 8.818.1023 Instructions for Use

21

ENABLED

Enable program start (a disabled program is

not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are
displayed if program is wrong).


List of program instructions

The following program instructions can be added to the time program

on the

Program

page:

Befehl Parametereintrag Bedeutung

Pump on,

off

Switch the high pressure pump on or
off.

Flow

0.01 ... 5 mL/min

Set flow rate of the high pressure
pump to the desired value.

Pmax

0.0 ... 30.0 MPa

Set maximum pressure limit for the
high pressure pump to the desired
value.

Pmin

0.0 ... 30.0

MPa

Set minimum pressure limit for the
high pressure pump to the desired
value.

Links

Die Registerkarte

Links

im Fenster

818 IC Pump

dient zur Wahl und

Einstellung der COM-Schnittstelle (Details siehe section 5.2.4 Links of

the «IC Net 2.3» Instructions for Use 8.110.8283).