Metrohm 818 IC Pump User Manual

Page 26

Advertising
background image

3 Operation

818 IC Pump / 8.818.1023 Instructions for Use

20

switched off. At the same time, the

Pressure

field color in the 818 settings window is
changed to the value set for

Out of range

.

<Set>

Send current parameters immediately to the
818 IC Pump (only available if pump is
running). Parameters are not stored in the

system file (

*.smt

) as long as the file is not

saved.

<Start>

Set current parameters and start 818 IC Pump

drive.

<Stop>

Stop 818 IC Pump drive.



Time program

The

Program

tab of the

818 IC Pump

window a user-defined time program

can be entered. This program is started automatically as defined in the

Start mode

window (see section 4.4.3 «IC Net 2.3» Instructions for Use

8.110.8283), either at the moment the determination is started (

Start with

determination

) or at the moment the sample is injected (

Start with inject

).A

First column

Time at which program instruction is applied.

Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruction is
applied together with the last instruction with
time entry.

Second column

Program instruction (see below).

Third column

Parameter for program instruction (see below).