3 Operation

818 IC Pump / 8.818.1023 Instructions for Use

22

Advanced

The

Advanced

tab of the

818 IC Pump

window is only available for a

connected system.

Correction factor

Flow correction factor to adjust the displayed
flow to the actual flow measured with the aid of
a measuring cylinder.
Entry range:

0.90 ... 1.10

Correction factor =

Displayed flow rate (in mL/min)

Measured flow rate (in mL/min)

<Set>

Send

current

Correction factor

value to the 818

IC Pump. This parameter is not stored in the
system file (

*.smt

) as long as the file is not

saved.

The difference between displayed and actual flow rate can be

maximum 3 % owing to unavoidable tolerances in the manufacture of

mechanical parts. Examine therefore the flow rate after

disassembling/assembling of the pump head and adjust the 818 IC

Pump with the correction factor if necessary.