3.3 Settings in the "818 IC Pump" window

818 IC Pump / 8.818.1023 Instructions for Use

19

Settings

The

Settings

tab of the

818 IC Pump

window contains basic settings of

the 818 IC Pump.

Pressure

Live display of current pressure in MPa.

Flow

Live display of current flow in mL/min.

New flow

A new flow can be set by moving the scroll bar,
by entering a value or by changing the value

using the up and down arrows.
Entry range:

0.01 ... 5.00 mL/min

Min/Max pressure

The

minimum and maximum pressure limit for

the 818 IC Pump can be set in steps of 0.1
MPa.
Entry range:

0.0 ... 30.0 MPa

.

The

set

maximum limit value should lie

between 5 to 10 MPa above the particular
operating pressure or the maximum admissible
operating pressure of the column used. If the
pump exceeds the preset limit value during

operation, it is switched off immediately. At the
same time, the

Pressure

field color in the 818

settings window is changed to the value set for

Out of range

.

The

set

minimum limit value should lie far

enough below the particular operating
pressure. If the pump pressure falls below this
preset lower limit during operation and this

pressure drop persist for several revolutions of
the pump cam due to leaks or interrupted inflow
of the eluent, the pump is automatically