2.7 Software installation

818 IC Pump / 8.818.1023 Instructions for Use

17

2.7 Software

installation

The PC program «IC Net 2.3» is required for the operation of the 818 IC

Pump. This program runs under Windows 2000 and Windows XP

operating systems and is installed according to section 1.4.2 of the

Instructions for Use 8.110.8283 from the Metrodata software «IC Net

2.3».
The installation of 818 IC Pump is described in section 6.1.2 of the

Instructions for Use 8.110.8283 from the Metrodata software «IC Net

2.3». Take care that the RS 232 settings are identical for the RS 232

interface of 818 IC Pump and 830 IC Interface. For the settings of 818

IC Pump see section 6.10.

2.8 Deaerating

the

pump

1 Start IC system

Switch the IC instruments on, start «IC Net» and open the

device window

818 IC Pump

.

Operation of 818 IC Pump via the software «

IC Net 2.3» is described

more detailed in section 3. For further information refer to the

Instructions for Use of «

IC Net» and to the online help of the program.

2 Purge pump

The purge valve 2 is opened by turning anticlockwise and the

pump can now be deaerated by starting the pump in «IC Net».

The solvent bottle should be located somewhat above the pump, but

at least at the same level as the pump head

8

(never below the

pump).

3 Deaerate pump

Should the pump not prime by itself, the 6.2816.020 syringe

supplied is used to siphon off solvent via the outlet capillary 5

with opened purge valve through the pump head until there are

no more bubbles in the solvent entering the syringe. The pump

can then again be briefly purged.

4 Close purge-valve

On completion of this procedure, the purge valve 2 is closed by

turning clockwise. The pump is now ready for operation.