5 interfaces, 1 rs 232 interface, Interfaces – Metrohm 818 IC Pump User Manual

Page 37: Rs 232 interface, Data transmission protocol, N 5 interfaces

Advertising
background image

5.1 RS 232 interface

818 IC Pump / 8.818.1023 Instructions for Use

31

5 Interfaces

5.1

RS 232 interface

5.1.1

Data transmission protocol

The RS 232 interface of the 818 IC Pump is configured as DTE (Data

Terminal Equipment) with the following technical specifications:

Standard

Data interface in accordance with EIA standard

RS 232C (DIN 66020, page 1)

Baud rate 9600

Data Bits

8

Stopp Bits

1

XON/XOFF

off

Control characters C

R

DEC 13

HEX 0D

L

F

DEC 10

HEX 0A

Max. line length

80 characters + C

R

L

F

Cable

For interconnections of the 818 IC Pump with

non-Metrohm units, only a shielded data cable

(e.g. Metrohm D.104.0201) may be used. The

cable shielding must be faultlessly earthed at

both units (pay attention to current loops;

always use star-head earthing). Only

connectors with adequate shielding may be

used (e.g. Metrohm K.210.0001 with

K.210.9004).

Cable length

max. ca. 15 m