Metrohm 865 Dosimat plus User Manual

Page 42

Advertising
background image

5.2 Managing solutions

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

865 Dosimat plus

Cylinder volume

Cylinder volume of the buret unit in mL. The cylinder volume is automati-
cally read out with intelligent buret units.

Selection

1 | 5 | 10 | 20 | 50

Default value

20

Concentration

Concentration of the solution.

Range

–999999999 ... 9999999999

Default value

1.000

Concentration unit

Unit of the concentration.

Selection

µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L | User-defined

Default value

mol/L

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.

Titer

Titer of the solution.

Range

–999999999 ... 9999999999

Default value

1.000

Titer unit

Unit of the titer.

Selection

µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L | empty | User-defined

Default value

empty

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.

Date titer det.

Date of the last titer determination.

Monitoring

Activating and deactivating the titer monitoring.

Advertising