6.3.2

Selecting a solution

Menu

Parameters

Solution

Selection of the solution from the solution list. We recommend always
selecting the solution. Solutions are defined under System

Solutions.

For exchange units with integrated data chip, a check is made in the
method run to verify whether the correct solution has been attached. For
exchange units without integrated data chip, only the cylinder volume is
checked. The validity of the titer is checked for the selected solution at the
start of the determination.

Selection

Selection of configured solutions | not defined

Default value

not defined

not defined
No check takes place.

6.3.3

Definition of the target solution

Menu

Parameters Content definitions

Content

Target concentration of the solution.

Input

10 characters

Range

0.000000001 ... 9999999999

Default value

1.0

Content unit

Unit of the target concentration.

The unit of the target concentration determines the calculation formula
for determining the volume to be dosed.

Selection

% | g/L | mg/L | mmol/L | mmol/kg | mol/L |
mol/kg | ppm

Default value

mol/L

Molar mass

Molar mass of the starting substance.

If selected unit is: mmol/L, mmol/kg, mol/L or mol/kg.
Input

10 digits

Range

0.000000001 ... 9999999999 g/mol

Default value

1.0 g/mol