1 commands and variables, Commands and variables, Figure 26 – Metrohm 865 Dosimat plus User Manual

Page 81: Connecting the rs-232/usb box to the pc

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Appendix

865 Dosimat plus

■■■■■■■■

71

6.2134.040

Figure 26

Connecting the RS-232/USB Box to the PC

Interface parameters
We recommend the following parameters for the RS-232 interface
COM2:

Baud rate: 19200

Data bits: 8

Stop bits: 1

Parity: none

Handshake: software

On the 865 Dosimat plus the interface parameters are set under
Menu

System External devices COM2 settings.

Transfer protocol
The data communication is synchronous. On each command there is an
answer by the instrument.

A command must be sent to the instrument with the control character CR
LF
as terminator. The answers of the instrument will also be transferred
with CR LF as terminator.

The instrument does not send spontaneous messages.

8.8.1

Commands and variables

Command

Function

Comment

$G

Start/Continue

Corresponds to the key [START] or [Con-
tinue]

$S

Stop

Corresponds to the key [STOP]

$H

Hold

Hold the method run