6 Parameters and dosing modes

865 Dosimat plus

■■■■■■■■

53

6.3.5

Reports

Menu

Parameters Reports

The reports that will be printed out automatically in connection with a
determination are defined under Reports.

Results

The result report contains the calculated result and additional specifica-
tions.

Selection

Label 1 | off | on

Default value

off

Label 1
A result report is printed out in a simple label format.

Example of a label:

—————————————————————————————————————————
Solution Method0012
—————————————————————————————————————————

Content 1.000 mol/L
Prepared at: 08-02-05 14:15:01
Prepared by: User1

Expires: ______________________

PC/LIMS

The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 34)
.

The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt
.