Metrohm 865 Dosimat plus User Manual
Page 60
6.3 Creating solutions (CNT D)
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
865 Dosimat plus
where:
= dosed volume in ml
= sample size in g
= optional factor, e.g. for non-100 % samples
= conversion factor, depending on unit
= target concentration in the selected unit
= molar mass in g/mol
= density of the dosed solvent in g/ml
Table 1
Variables and constant values in the CNT D mode
Type of calculation
Unit
Substance concentration
mol/L
10
3
.
.
1
Substance concentration
mmol/L
10
6
.
.
1
Mass concentration
g/L
10
3
.
1
1
Mass concentration
mg/L
10
6
.
1
1
Mass fraction
%
10
2
.
1
.
Mass fraction
ppm
10
6
.
1
.
Molality
mol/kg
10
3
.
.
.
Molality
mmol/kg
10
6
.
.
.
.
= input possible (default 1)
Performing a dosing
Pressing the [GO] key causes the dosing to run automatically. The sample
data is requested first. The calculated target volume is displayed. The dos-
ing is started after the [GO] key has been pressed once again.
The dosing can be canceled with the [FILL] key. The dosing cylinder is fil-
led automatically.