5 system initialization, System initialization – Metrohm 865 Dosimat plus User Manual

Page 75

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Appendix

865 Dosimat plus

■■■■■■■■

65

Note

Inexpensive printers are often designed solely for use with a PC and
may not be equipped with one of the printer languages listed above.
Such models are not suitable for this reason.

8.5

System initialization

In very rare instances, it could happen that a faulty file system (e.g.
because of a program crash) will lead to an impairment of program func-
tioning. The internal file system needs to be initialized in such cases.

Caution

All user data (methods, solutions, etc.) are deleted if a system initializa-
tion is carried out. Afterwards, the instrument will have the factory set-
tings again.

We recommend creating a backup copy of the system at regular inter-
vals in order to avoid data losses.

After a system initialization the program versions and language files do
not have to be reloaded. Only the selection of the dialog language may
have to be reset in the system settings.

Proceed as follows for system initialization:

1 Switch off the instrument

Keep the red [FILL] key pressed down for at least 3 s.

A progress bar is displayed. If the key is released during this time,
then the instrument will not be switched off.

2 Switch on the instrument

Keep the red [FILL] key pressed down for approx. 10 s.

The dialog for confirmation of the initialization is displayed for 8 s.
The initialization must be confirmed during this time.