6.1 Manual dosing (DOS)

■■■■■■■■■■■■■■■■■■■■■■

42

■■■■■■■■

865 Dosimat plus

Note

If a solution has been selected under Parameters, then the corre-
sponding titer and the concentration will be read out from the solution
data under System

Solutions and used for calculations. If this is not

the case, then calculations will be made with the Default value 1.

Calculation status

Activating the calculation.

Selection

on | off

Default value

off

on
The settings for the calculation can be entered as soon as the calcula-
tion has been activated.

Result name

The result name will be shown in the result display and in the report.

Input

12 characters

Decimal places

Number of decimal places used to display the result.

Range

0 ... 5

Default value

2

Result unit

The result unit is displayed and saved together with the result.

Proceed as follows if you wish to delete a selected unit:

Delete the unit entered under User-defined. A blank entry will be gener-
ated in the selection list.

Selection

% | /pc | L | g | g/L | mL | mg | mg/mL | mol |
mol/L | ppm | User-defined

Default value

%

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined.

Factor

Factor for the calculation formula.