6 Parameters and dosing modes

865 Dosimat plus

■■■■■■■■

55

The run can be canceled with the [FILL] key. The dosing cylinder is filled
automatically.

Up to five steps are carried out. A text is displayed on the instrument as a
demand note with each step. The readiness for carrying out the next step
is indicated by the green light of the LED on the Manual Dosing Control-
ler. Advance to the next step with the green [GO] key. If no Safety air
gap
or Diluting volume has been selected (0 mL), then the correspond-
ing steps will not be carried out.

Display

Step 1 (of 5)

Hold tip into waste vessel.

Hold the pipetting tip in a waste vessel and press [Go].

The pipetting volume will be expelled in order to make
room in the pipetting tubing.

Step 2 (of 5)

Lift tip above surface for aspi-
ration of air gap.

Raise the pipetting tip and press [Go].

The air gap will be aspirated.

Step 3 (of 5)

Immerse tip into sample.

Immerse the pipetting tip in the sample and press [Go].

The sample will be aspirated.

Step 4 (of 5)

Lift tip above surface for
safety air gap.

Raise the pipetting tip and press [Go].

The air gap will be aspirated.

Step 5 (of 5)

Hold tip into target vessel.
Sample will be ejected.

Hold the pipetting tip in an empty vessel and press
[Go].

The sample will be expelled.

The dilution volume will be expelled.

Solutions and buret units
Buret units with or without an integrated data chip can be used for dos-
ing.