8.8 Remote control via an RS-232 connection

■■■■■■■■■■■■■■■■■■■■■■

72

■■■■■■■■

865 Dosimat plus

Command

Function

Comment

$D

Scan instrument sta-
tus

Answers: Ready;0 , Busy;0 , or Hold;0 (0
= no message)

If a message on the instrument requires
the interaction of the user, the answer of
the status scan displays the corresponding
message number. Example:

Busy;010-119

= "Check buret unit"

The message can be quit with [OK] or
[Cancel], see below.

$A

Quit message

Confirm the message on the instrument
with [OK]

A mandatory status scan providing the
message number must take place immedi-
ately before quitting the message, see
above.

$A(OK), $A(CANCEL)

Quit message

Confirm the message with [OK] or [Can-
cel]

$A(YES), $A(NO)

Quit message

Confirm the message with [Yes] or [No]

$L(method name)

Load method

The method name has to be known and
unique.

$Q(variable)

Request variable value

List of variables, see below

Variable

Comment

VOLUME

Dosed volume

TITER

Titer of selected solution

CONC

Concentration of selected solution

C00

Sample size (DOS only)

RESULT

Result (DOS only)

RATE

Dosing rate (XDOS only)

TIME

Dosing time (XDOS only)

CONTENT

Content (CNT only)

The values of the variables are only available after the end of a determina-
tion (in the status 'ready').