Density

Density of the solvent to be dosed.

If selecting unit is: %, mmol/kg, mol/kg or ppm.
Input

10 digits

Range

0.000000001 ... 9999999999 g/mL

Default value

1.0 g/mL

Factor

Multiplication factor for the calculation of the volume to be dosed. The
factor can be used in general as a correction factor, e.g. for compensating
for volume contraction or for specifying the starting concentration.

Input

10 digits

Range

0.000000001 ... 9999999999

Default value

1.0

6.3.4

Dosing parameters

Menu

Parameters Dosing parameters

Dosing rate

The rate at which dosing takes place. The maximum dosing rate depends
on the cylinder volume (see Chapter 8.1.1, page 61).

Range

0.01 ... 150.00 mL/min

Selection

max.

Default value

max.

Filling rate

Rate at which the dosing cylinder is to be refilled. The maximum filling
rate depends on the cylinder volume (see Chapter 8.1.1, page 61).

Range

0.01 ... 150.00 mL/min

Selection

max.

Default value

max.

Volume limit

Safety limit for limiting the maximum volume to be dosed.

Range

0.01 ... 99999.9 mL

Selection

off

Default value

off

off
No safety limit.