6 Parameters and dosing modes

865 Dosimat plus

■■■■■■■■

47

6.2.4

Dosing parameters

Menu

Parameters Dosing parameters

Dosing criteria

Selection of the preset dosing criteria.

Selection

Rate/Time | Volume/Rate | Volume/Time

Default value

Volume/Rate

Rate/Time
Dosing is carried out at the selected dosing rate for the amount of time
specified. The time is also running while the cylinder is being filled.
Volume/Rate
The volume specified is dosed at the selected dosing rate.
Volume/Time
The volume specified is dosed for the time period selected. The dosing
rate necessary for this, is calculated according to these specifications.
The time needed for filling the dosing cylinder and for switching over
the flat stopcock is taken into account thereby.

Volume

The volume to be dosed.

Range

0.0000 ... 99999.9 mL

Default value

10.0000 mL

Dosing rate

The rate at which dosing takes place. The maximum dosing rate depends
on the cylinder volume (see Chapter 8.1.1, page 61).

Range

0.01 ... 150.00 mL/min

Selection

max.

Default value

max.

Filling rate

Rate at which the dosing cylinder is to be refilled. The maximum filling
rate depends on the cylinder volume (see Chapter 8.1.1, page 61).

Range

0.01 ... 150.00 mL/min

Selection

max.

Default value

max.

Volume limit

Safety limit for limiting the maximum volume to be dosed.

Range

0.01 ... 99999.9 mL