background image

1 Introduction

826/827 pH meter, Instructions for Use

3

Fig. 3: Rear view of the 827 pH lab

4 Connection for potentiometric elec-

trodes

pH, ISE, redox or silver electrodes

with built-in or separate reference elec-

trode; socket type F

7 Infrared interface

Connection for printer or PC with infra-

red interface

5 Connection for separate reference

electrode

for connection with two 2 mm B-plugs;

the adapter 6.2103.180 or 6.2103.190

is necessary with 4 mm banana plugs

8 6 V power connection

with 827 pH lab only

6 Connection for temperature sensor

Pt1000 or NTC, for connection with two

2 mm B-plugs; the adapters

6.2103.180 and 6.2103.190 are neces-

sary with 4 mm banana plugs; red plug

in socket "Temp."!