4 maintenance, 1 changing the batteries (826 only), Maintenance – Metrohm 827 pH lab User Manual

Page 57: Changing the batteries (826 only)

Advertising
background image

8 Troubleshooting – Messages – Maintenance

826/827 pH meter, Instructions for Use

49

8.4

Maintenance

Warning!

You should remove the batteries in case the pH meter is not used for

a longer period. Leaking batteries can damage the pH meter.

8.4.1

Changing the batteries (826 only)

The 826 pH mobile reminds you of changing the batteries by displaying

the message "battery low" and the symbol . The measurement ac-

curacy is not affected at all.

Note!

All the parameters are reset to the default values. The current date

and time are reset to 2005-01-01 and 00:00:00 respectively. The cali-

bration data as well as the stored measured values are not lost.
Metrohm recommends using batteries of the type alkaline or lithium.

Proceed as follows:

Switch off the pH meter.

Unscrew the six screws on the base of the housing with the en-

closed allen key and remove the base plate.

Fig. 5: Changing the batteries for the 826 pH mobile

Warning!

Do not remove the gray shielding plate under any circumstances and

do not touch any electronic components.