4.2 pH calibration

18

826/827 pH meter, Instructions for Use

3 Result

Wait for potential measurement, the result will be displayed:

After 30 s the pH meter will switch back automatically to the

measurement display. This can also be done immediately with

<OK> or <QUIT>.

3 Poor calibration data

If the calibration data are outside the calibration parameters

defined as the limits (see Section 6.1.3) then a corresponding

message will be shown:

You can still accept the calibration data with <OK>, or reject

it with <QUIT>.

4 Show calibration data

You can now display the calibration data under param/

cal.data.

5 Print calibration report

Change to the menu << report >> with the arrow keys <>

or <>.

Press <OK>.

Select the report cal with the arrow keys <> or <> and

start the printing with <OK>.