9 system test report, System test report – Metrohm 827 pH lab User Manual

Page 47

Advertising
background image

7 Various functions

826/827 pH meter, Instructions for Use

39

7.1.9

System test report

After the pH meter has been switched on an electronic system test is

carried out automatically.
The tests are documented in the system test report which can be

printed out when the pH meter is switched on: (see Section 5.1):

'di

827 pH lab

01106 5.827.0010

print date

2005-01-28 08:16:41

EPROM test OK

ADC adjust test OK

real time clock OK

A/D converter OK

LCD display OK

==============

Note!

The system test report can't be printed manually via the menu << re-

port >>.

If one of these tests is not concluded with 'OK' then please contact Me-

trohm Service.
With a real time clock error you can try resetting the date and time. If

the test is then 'OK', you should check whether the instrument configu-

ration is still unchanged.