8 Troubleshooting – Messages – Maintenance

826/827 pH meter, Instructions for Use

43

8

Troubleshooting – Messages –

Maintenance

8.1

Troubleshooting

If problems occur during pH measurement then these can have various

causes:
1. Application

Difficult sample matrices or interferences can prevent reliable mea-

surements (e. g. insufficient ionic strength). Our Application Bulle-

tins and Application Notes support you in correct selection of the

analytical conditions. The most suitable electrode for your applica-

tion can be found under www.metrohm.com.

2. Buffer solution

The precision of pH measurements depends primarily on the correct

calibration of the electrode. You should use clean and fresh buffer

solutions. For example, a frequent cause of incorrect calibration is

the use of an old pH 9 or pH 10 buffer, whose pH can vary consi-

derably from the certified pH of a new buffer as a result of the ab-

sorption of atmospheric CO

2

.

3. Measuring / Reference electrode

The electrodes are the most important element of the whole measur-

ing system. The correct handling of the measuring and reference

electrodes is described in the leaflet which accompanies the elec-

trode.

4. pH meter

If the pH meter is thought to be the cause of a measuring problem

then first check all the configuration and parameter settings. The di-

agnostic functions help you in searching for the fault (see Sec-

tion 8.3).

The pH meter will inform you directly about problems during the mea-

surement. These messages are explained in the following Section 8.2.

The following table lists general problems that can occur during pH

measurements. Possible causes and appropriate remedies are also

described.