4 Short operating tutorial

826/827 pH meter, Instructions for Use

17

4

Short operating tutorial

In this section the necessary steps for carrying out a simple pH and ion

measurement with calibration are described. The instructions are li-

mited to those steps that are absolutely necessary and will enable you

to carry out your first measurements with the pH meter directly. The op-

erating principles are described in Section 3.3.

4.1

Requirements

The following instruments, accessories and solutions are required for

carrying out the pH calibration and measurement described below:

826 pH mobile (2.826.0XX0) or 827 pH lab (2.827.011X,

2.827.021X)

pH electrode

Calibration buffers

Metrohm buffer solutions pH 4.00 and pH 7.00

The calibration parameters for the pH mode are set for a calibration us-

ing two Metrohm buffers as default (see Section 6.1.2). If you want to

use other buffers then the corresponding buffer type must be entered.

4.2

pH calibration

1 Start calibration with first buffer

Start calibration with <CAL>

Immerse pH electrode in buffer solution pH 7 and confirm with

<OK>

If a temperature sensor is connected, the calibration tempera-

ture is being determined

If no temperature sensor is connected:

enter the temperature with the arrow keys and confirm with

<OK>

The first buffer is being measured

The following message is displayed after successful mea-

surement:

change buffer <OK>

2 Continue calibration with second buffer

Remove pH electrode from first buffer and rinse with water

Immerse pH electrode in second buffer solution pH 4 and

continue calibration sequence with <OK>

The second buffer is being measured