6.1 pH measurement (pH mode)

30

826/827 pH meter, Instructions for Use

6.1.2

Calibration parameters

param

:

cal.settings

: ├

temperature

report

cal.interval

no.of buffers

buff.type

The calibration parameters describe the most important settings for the

calibration such as buffer types and number of buffers.

temperature

0.0...25.0...99.9 °C

Just as for the measuring temperature, the calibration temperature

should also be entered when no temperature sensor is connected.

Otherwise the temperature is measured continuously and this value will

be ignored. If the measuring and calibration temperatures are different,

then this is absolutely necessary for the automatic temperature correc-

tion of the electrode slope. The temperature is also stored with the suf-

fix 'manual' and marked in this way in the calibration report.
The exact temperature recording is also important for the pH calibration

as the pH meter works with automatic buffer recognition (see below:

buff.type).

report:

on, off

After the calibration it is possible to output the calibration data automat-

ically as a report.
Such a report can also be printed out at a later date.

cal.interval

0...999 h, 0 = off

The pH meter can automatically remind you about any recalibration of

the pH electrode used. When this interval has elapsed, a message will

appear on the display:

cal.interval exp.

This message can be deleted from the display with <QUIT>. It is dis-

played again every 10 min until a new calibration has been carried out.

Until then it will be printed out on every measuring point report.

Note!

The calibration interval is only valid after the next calibration.