5 Configuration

826/827 pH meter, Instructions for Use

25

5.3

Storing measured values

<< config >>

configuration

:

store

: └

store crit.

immed., drift, off

time

time int.

stop time

change

delta pH/T/mV

stop time

Up to 200 measured values can be stored manually or automatically af-

ter reaching the drift criterion, at definite time intervals or after changing

by a certain amount (see Section 7.2). The storing is done by pressing

<OK> in the menu << store >>.

store crit.:

immed., time, drift, change, off

Depending upon the store criterion a measured value will be stored ei-

ther immediately, at defined time intervals, when the drift condition

has been fulfilled or after a certain difference of the measured value

has been exceeded.
The following additional information are stored with each measured val-

ue:

Temperature (Mode pH only)

Id

Date

Time

store crit.:

immed.

The current measured value is stored with <OK> if you are in the

menu << store >>.

store crit.:

time

time int.:

1...4...99'999 s

stop time:

0...999'999 s

You can automatically record and save the measured values at fixed

time intervals until the stop time has been reached.

store crit.:

drift

A measured value can only be stored if the message drifting... is

no longer displayed.

store crit.:

change

The next measured value is stored automatically not until the difference

to the previous measured value is bigger than the parameter delta

pH/T/mV defined here.