3 ph measurement, Ph measurement – Metrohm 827 pH lab User Manual

Page 27

Advertising
background image

4 Short operating tutorial

826/827 pH meter, Instructions for Use

19

6 Possible error messages

In case buffer pH 7 has been measured again by mistake, the

following message is displayed:

same buffer

Change the buffer and press <OK>. The calibration is being

continued. The calibration can be canceled with <QUIT>.

If the temperature difference is > 2 °C between the buffers,

the following message is displayed:

delta T > 2 °C

Adjust the temperature and continue the calibration with

<OK> or cancel it with <QUIT>

4.3

pH measurement

1 Select print criterion

If the obtained measured value is to be printed out directly as

a measuring point report then the required print criterion must

be set (see Section 5.2):

configuration/print crit.: drift

2 Start measurement

Immerse the pH electrode in the sample

3 Determine measured value

You can read off the current pH value of the sample from the

display. It is stable when the message drifting... is no

longer being displayed (preset drift threshold: 0.028 pH/min)

1

4

.

Print measured value

Change to the menu << print >> with the arrow keys <>

or <>.

The output of a measuring point report on a printer is started

with <OK>. Due to the print criterion drift the recording of

the measurement and its output takes place only when the

measured value is stable:

1

Further preset drift thresholds:

for potential measurement: U/mV: 1.875 mV/min, for temperature measurement: t/°C: 0.974 °C/min

'mp
827 pH lab 01106 5.827.0010
print date 2005-02-08 15:43:01
pH = 4.612 22.8 °C NTC
==============