2 Installation

826/827 pH meter, Instructions for Use

9

2.6

Connecting a printer

A printer with an infrared interface can be connected to the infrared in-

terface for printing out reports.
Activate the infrared interface of the pH meter under
configuration/auxil/IR interface: (see Section 5.4)
No further transmission parameters need to be set.

Attention!

In order to guarantee a perfect transmission the distance between the

pH meter and the printer should be 80 cm at the maximum. The infra-

red interfaces must face each other directly. Never expose the infra-

red interface to direct sunlight.

2.7

Initial configuration

The pH meter is delivered with standard settings for the configuration. If

it should ever be necessary to reset the configuration of the instrument

to its original condition then this can be carried out by a re-initialization

of the instrument memory (see Section 8.3).
Before you start to make measurements please change the following

configuration settings. If you first want to make yourself familiar with the

operation of the pH meter then please read Sections 3.1 to 3.3. More

detailed explanations of the individual configuration settings are given

in Section 5.

Date and time

configuration/auxil/date

and /time

Please check that the date and time are correct.

Temperature sensor

configuration/auxil/temp.sens.
Enter the type of temperature sensor that is connected. If no tempera-

ture sensor is connected and you always want to enter the measuring

temperature manually then you can ignore this setting.