5.2 Printing out measured values

22

826/827 pH meter, Instructions for Use

line feed

0...3...999

The paper feed of the printer at the end of a report can be set here. You

can select the number of empty lines so that after the report has been

printed out you can tear off the printed paper directly at a suitable point.

The paper feed depends on the type of printer and has to be set accor-

dingly.

Note!

The line feed must be set to 999 when using a A4 printer.

dev.label

8 ASCII characters

If you wish to assign the printed reports to a particular pH meter not on-

ly by the unchangeable device identification, you can enter any text

here. This entry will be printed out in each report provided that an identi-

fication has been entered.

sys.test rep.

on, off

After switching on the 826/827 pH meter a system diagnosis is carried

out automatically. If this parameter is switched on, the result is printed

automatically after switching on the pH meter.

5.2

Printing out measured values

<< config >>

configuration

:

print

: └

print crit.

immed.

date & time

rep.header

cal.report

time

date & time

rep.header

cal.report

time int.

stop time

drift

date & time

rep.header

cal.report

change

date & time

rep.header

cal.report

delta pH/T/mV

stop time

off

In the configuration menu print the output form of measured values,

triggered by <OK> in the menu << print >> is defined.

print crit.:

immed., time, drift, change, off

Depending upon the print criterion a measured value report will be

transmitted to a printer either immediately, at defined time intervals,