6.1 pH measurement (pH mode)

32

826/827 pH meter, Instructions for Use

The other stored buffer tables are described in Section 9.3. During the

calibration the pH meter shows the type and temperature-specific pH

value of the recognized buffer. For temperatures within the 5 °C steps

the pH value is linearly interpolated.

buff.type:

special

If you would like to use different buffers from those described,

then you require the exact pH value of each buffer at the tempera-

ture used for calibration. These pH values can be entered here.

During the calibration they can again be adapted before each

measurement.

A general rule is that you can only calibrate in defined temperature

ranges, as otherwise the following error message appears:

buffer not defined

6.1.3

Limits

param

:

cal.limits

: ├

lo.lim.slope

up lim.slope

lo.lim.pH(0)

up lim.pH(0)

The most important calibration data such as slope and pH(0) must be

within the limits defined here in order to be adopted automatically.

lo.lim.slope

1.0...95.0...999.9 %

up lim.slope

1.0...103.0...999.9 %

lo.lim.pH(0)

0.00...6.40...99.99

up lim.pH(0)

0.00...8.00...99.99

If one of the values of the calibration data lies outside these limits then

at the end of the calibration the following message is displayed:

cal. out of limits

You can accept these calibration data nevertheless with <OK> or you

can reject them with <QUIT>.