Appendix

In this section you will find the most important technical data of the pH

meter, dialog structures, a list of standard accessories and optional ac-

cessories as well as warranty and declarations of conformity.

9.1

Technical data

Provided that nothing to the contrary is mentioned, the published val-

ues represent the typical data of the 826 pH mobile and the 827 pH

lab.

9.1.1

Measuring modes

Measuring mode

Prim. measured

quantity

Sec. measured

quantity

pH

pH

T

Temperature

T

Potential

U

9.1.2

Measuring inputs

Potentiometric
for pH value, potential
- 1 high-impedance measuring input for pH, redox and ISE electrodes
- 1 reference input for separate reference electrode

Input resistance

> 1 * 10

12

Ohm (under reference conditions)

Temperature
Also for automatic temperature compensation
- 1 measuring input for temperature sensors (Pt1000 or NTC)
NTC characteristics configurable

Default values

R (25°C) = 30'000 Ohm / B

25/50

= 4100

Measuring interval

Measuring cycle

1 s for all measuring modes