3 attaching the accessories, 1 capillaries, 2 peek connectors – Metrohm 709 IC Pump User Manual

Page 17

Advertising
background image

2.3 Attaching the accessories

709 IC Pump

9

2.3

Attaching the accessories

2.3.1 Capillaries

Some of the connections under high pressure between the 709 IC

Pump and the IC system must be set up by the user. For the pressure

range 0

25 MPa (0

250 bar) the 6.1831.010 PEEK capillary

(i.d. = 0.25 mm, e.d. =

1

/

16

"

, length = 3 m) supplied in the accessories

of the 709 IC Pump can be used for both pump versions. For the pres-

sure range 25

50 MPa (250

500 bar), which is permissible only for

the not metal-free 2.709.0010 version, the 6.2620.020 Steel capillary

(i.d. = 0.25 mm, e.d. =

1

/

16

"

, length = 3 m) must be used.

With PEEK capillaries the connection is made preferably with the

6.2744.010 PEEK connectors (see section 2.3.2), with steel capillaries,

the 6.2620.000 and 6.2620.010 steel connectors should be used (see

section 2.3.3).

Capillaries fitted with new connectors must have a perfectly flat cut

surface. For PEEK capillaries it is best to use the

6.2621.080 Capil-

lary tubing cutter, for steel capillaries the 6.2621.040 Capillary

tubing cutter.



41 Ferrule

(6.2620.010)

42 Pressure screw (6.2620.000)

43 Capillary

6.2620.020 Steel capillary or

6.1831.010 PEEK capillary

44 Compression fitting

(6.2744.010)


2.3.2 PEEK

connectors

For the connection of 6.1831.010 PEEK capillaries or of 6.1822.010

PTFE Microcapillaries (i.d. = 0.3 mm), the supplied 6.2744.010 PEEK

Compression fittings are used. It is best to proceed as follows:

1 Mount compression fitting

Slide a compression fitting 44 (6.2744.010) over the end of the

capillary 43 to be fastened as shown in Fig. 3.

42

41

43

44

Fig. 3:

Connectors for capillaries

Advertising