7 remote control commands – Metrohm 709 IC Pump User Manual

Page 50

Advertising
background image

5 Interfaces

709 IC Pump

42

5.1.7

Remote control commands

Object Meaning

Entry

range

&

Root

.

Pump

Pump

.

Start

Start the pump

$G, $S

.

Soft

Soft start the pump (within 5 s)

$G, $S

.

Identifier

Instrument identification

up to 8 ASCII characters

.

Parameter

Parameters

.

A

Metrohm pump head

.

Flow

Flow rate

0.01

5.00 mL/min

.

Pmax

Upper shutoff pressure

P

min

50.0 MPa

.

Pmin

Lower shutoff pressure

0.1

P

max

MPa

.

NX

Correction factor

0.90...1.10

.

P

Parameter for prep. head

.

Flow

Flow rate

read only

.

Pmax

Upper shutoff pressure

read only

.

Pmin

Lower shutoff pressure

read only

.

NX

Correction factor

read only

.

M

Parameter for micro head

.

Flow

Flow rate

read only

.

Prnax

Upper shutoff pressure

read only

.

Pmin

Lower shutoff pressure

read only

.

NX

Correction factor

read only

.

R

Parameter for reserve head

.

Flow

Flow rate

0.01...5.00 mL/min

.

Pmax

Upper shutoff pressure

0.1

50.0 MPa

.

Pmin

Lower shutoff pressure

0.1

50.0 MPa

.

NX

Correction factor

0.90

1.10

.

Info

Current configuration

.

Head

Pump head

.

Actual

Current data

.

Pressure

Pressure

.

Flow

Flow rate

.

Pmax

Upper shutoff pressure

.

Pmin

Lower shutoff pressure

.

NX

Correction factor

.

MaxPmax

Maximum possible upper shutoff pressure

.

MinPmax

Minimum possible upper shutoff pressure

.

MaxPmin

Maximum possible lower shutoff pressure

.

MinPmin

Minimum possible lower shutoff pressure

.

Setup

Instrument settings

.

Trace

Message if values or paths change

ON,OFF

.

Autoinfo

Automatic message on changes:

.

P

when power switched on

ON,OFF

.

R

when instrument in "ready" status

ON,OFF

.

S

when instrument in "stop" status

ON,OFF

.

E

for error message

ON,OFF

.

Pressure

on every change of measured pressure value

ON,OFF