11 setting the baud rate – Metrohm 709 IC Pump User Manual

Page 53

Advertising
background image

5.1 RS232 interface

709 IC Pump

45

5.1.11 Setting the baud rate

The baud rate of the RS232 interface of the 709 IC Pump is set to 9600

as standard. If you wish to set a lower baud rate, proceed as follows:

The electronic circuitry of the 709 IC Pump is sensitive to electrostatic

charges. You should thus take the appropriate measures (earthing,

e.g. by touching metal) before opening the instrument.

1 Disconnect cables

Disconnect mains cable from mains connection plug 31 of the

709 IC Pump, also unplug all other cables connected to the rear

of the instrument.

2 Remove housing cover

Unscrew the two fastening screws 25 at the bottom of the

instrument and the two fastening screws 26 at the rear of the

instrument, remove housing cover 27.

3 Set baud rate

Change over blue jumpers at positions P21, P22 and P23 of the

main board (on right side of instrument when viewed from back)

using tweezers or flat-nosed pliers to match the baud rate you

wish:

P21
P22
P23

3 2 1

P21
P22
P23

3 2 1

P21
P22
P23

3 2 1

P21
P22
P23

3 2 1

P21
P22
P23

3 2 1

P2
P2
P2

3 2 1

9600 4800 2400

1200 600 300

4 Replace housing cover

Replace housing cover 27 and fasten with fastening screws 25

and 26.

5 Plug in cables

Connect mains cable to mains connection plug 31 and also plug

in all other cables.

Advertising