3 Operation

709 IC Pump

24

P

max

Maximum shutoff pressure for automatic

shutdown of the pump drive

Range: 0.1...50 MPa (1...500 bar)

(P

max

> P

min

)

P

actual

Current

pressure

NX

Correction factor for flow rate (see

section 3.2). The <SELECT> key must

be pressed for ca. 3 s.

Range: 0.90 ... 1.10

Input key for continuous lowering of the value selec-

ted with the <SELECT> key. When the key is

released the current value is stored.
Pressing this key briefly changes the displayed value

by one digit, if the key is pressed longer the value is

changed continuously and increasingly faster.

Input key for continuous raising of the value selected

with the <SELECT> key. When the key is released

the current value is stored.
Pressing this key briefly changes the displayed value

by one digit, if the key is pressed longer the value is

changed continuously and increasingly faster.

EXT.

On/off switching of the external control of the pump

via RS232 or current loop interface.

PURGE

On/off switching of the maximum delivery rate

(5 mL/min) to purge the pump.
This key is active only in the standby mode.