4 Notes – Maintenance – Faults

709 IC Pump

32

4.4 Diagnosis

4.4.1 General

information

The 709 IC Pump is a very precise and reliable instrument. Thanks to its

rugged construction it is virtually impossible for external mechanical or

electrical influences to have an adverse effect on its functions.
Although the occasional fault in the instrument can not be excluded

completely, it is certainly much more likely that malfunctions are caused

by wrong operation or handling or through improper connections and

operation with non-Metrohm instruments.
It is thus advisable in each case to isolate the fault with the rapid and

easy to perform diagnostic tests. The customer thus need not call

Metrohm service until there is a true fault in the instrument. In addition,

with the aid of the numbering in the diagnostic program he can provide

the service engineer with much more accurate information.
In inquiries always quote the manufacturing number 39 of the 709 IC

Pump (see Fig. 2), program version (see section 4.4.2) and specify

possible error messages.

Procedure

The diagnostic steps must be performed in sequence and compared

with the reactions of the 709 IC Pump (indented). In the "yes" case, con-

tinue with the next instruction.
If the instrument does not show the expected reaction ("no" case), the

appropriate diagnostic step must be repeated to exclude an operating

error. With repeated wrong reactions, however, there is a strong possi-

bility that a malfunction exists.

Equipment needed

3.496.8480 Test plug (Necessary only if external required functions

(RS 232) should also be checked).