2.3 Attaching the accessories

709 IC Pump

11

2.3.5

Filter unit PEEK

The 6.2821.100 Filter unit PEEK supplied (see Fig. 4) serves to avoid

contamination of the piston seals by abrasive particles and can be used

in the pressure range 0

25 MPa (0

250 bar). The filter unit consists of

the housing 46 and the two connectors 45 (with filter) and 47 (without

filter) to be screwed into the housing 46. For the connection of capillar-

ies 43, PEEK compression fittings 44 (6.2744.010) must be used.
New connectors 45 with filter are available as an option with the order-

ing number 6.2821.110 (10 pieces).

For the connection of the filter unit, please note the flow direction

arrow printed on the housing.

43

44

45

46

47

44

43

Fig. 4: Filter unit PEEK (6.2821.100)

43 Capillary

6.1831.010 PEEK capillary

46 Housing for filter unit

Part of 6.2824.100 Filter unit

44 Compression fitting (6.2744.010)

47

Connector without filter

Part of 6.2824.100 Filter unit

45 Connector with filter (6.2824.110)

Part of 6.2824.100 Filter unit