3.4 Pressure limit values for safety shutdown

709 IC Pump

21

3.4

Pressure limit values for safety shutdown

3.4.1

Minimum pressure limit

The minimum shutoff pressure of the pump can be set in steps of

0.1 MPa from 0.1

50 MPa with the pump drive switched on or off. Pro-

ceed as follows:

1 Select "P

min

"

Press <SELECT> key 4 until the "P

min

" LED 11 lights up.

2 Set shutoff pressure

Press <

> key 5 or <

> key 6 until the desired minimum

shutoff pressure (in MPa) appears in display 1.

3 Store shutoff pressure

Release <

> key 5 or <

> key 6: the value shown in display

1 at the time of release will be stored.

Entry of the value "0.0" switches off the lower pressure limit monitoring.

The set limit value should lie far enough below the particular operating

pressure. If the pump pressure falls below this preset lower limit during

operation, the microprocessor checks whether this violation persists for

several revolutions of the cam or whether it was a temporary drop in

pressure due to an air bubble or a brief leak in the valves. If the cause is

a temporary loss of pressure, the pump is not switched off. In the case

of a pressure drop due to leaks or interrupted inflow of the eluent, the

pump drive is switched off and disabled.

If the pump has been shut off by the minimum pressure monitoring, the

LED 11 flashes, the shutoff pressure can be requested in the display

"P

actual

" (see section 3.7). The safety shutdown is not active during the

first two minutes after the pump drive has been switched on with the

<R/S> key 3.
Pressing the <R/S> key 3 re-enables the pump drive, the LED 11 no

longer flashes and the pump is in the standby mode. Pressing the

<R/S> key 3 again puts the pump back into operation.