2 Installation

709 IC Pump

14

2.4.2

Connection to injection valve with steel capillaries

For the pressure range 25

50 MPa (250

500 bar), it is recommended

to use 6.2620.020 steel capillaries, a 6.2620.150 Pulsation dampener

(see section 2.3.4), and a Filter unit Manufit (see section 2.3.6) to con-

nect the not metal-free 2.709.0010 IC Pump to the injection valve of the

733 IC Separation Center. Proceed as follows:

1 Connection of filter unit Manufit

Attach inlet capillary 48 of the filter unit Manufit 58 to connec-

tion 23 of the 709 IC Pump (see Fig. 2).

Attach outlet capillary 55 of the filter unit Manufit 58 using a

coupling 59 to a steel capillary 43 cut to the required length

and equipped with connectors (see Fig. 7).

2 Installation of the capillary in the IC Separation

Center

Procedure, see 732/733 Instructions for Use, section 2.6.5.

3 Connection of the pulsation dampener

Procedure, see 732/733 Instructions for Use, section 2.6.5.

4 Connection to the injection valve

Procedure, see 732/733 Instructions for Use, section 2.6.5.



709

45

59

43

733

57

43

58 52

Fig. 7: Connection to injection valve with steel capillaries

43 Capillary

6.2620.020 Steel capillary

57 Pulsation dampener (6.2620.150)

48 Inlet capillary

of filter unit Manufit 58

58 Filter unit Manufit (6.2824.000)

55 Outlet capillary

of filter unit Manufit 58

59 Coupling (6.2620.060)