4 status messages – Metrohm 709 IC Pump User Manual

Page 48

Advertising
background image

5 Interfaces

709 IC Pump

40

5.1.4 Status

messages

To ensure appropriate control of an external control unit is possible, it

must also be possible to request status conditions which provide infor-

mation on the status of the 709 IC Pump. The output of a status mes-

sage is triggered by the trigger

'$D'

. Status messages consist of the

global status, the detailed status and eventual error messages, e.g.

'

$S.P.S;E32

'. The global status informs on the activity of the process,

while the detailed status conditions show the exact activity within the

process.

The following global status conditions are possible:

$G

Go

The pump is running.

$R

Ready

The pump is ready.

$S

Stop

The pump has been stopped by "unnatural means".

Detailed status conditions
Status conditions of the global $G:

$G.P.S

The pump has been started with normal start.

Status conditions of the global $R:

$R.P.S

The pump is ready, inactive.

Status conditions of the global $S:

The instrument gives the status from which it has been
stopped, e.g. $S.P.S;E32;

Advertising