2 control interface, 1 functions, 2 settings for the control interface – Metrohm 709 IC Pump User Manual

Page 54

Advertising
background image

5 Interfaces

709 IC Pump

46

5.2 Control

interface

5.2.1 Functions

The 709 IC Pump has the following control inputs and outputs, which

are accessible via the control interface:
Inputs

Reduce flow rate (possible via contact or TTL signal)

Stop pump (possible via contact or TTL signal)

Outputs

Signal when pump has been stopped or

Signal when pump is running

You will find detailed information on these functions in section 5.2.3,

possible changeover of the jumpers is described in section 5.2.2.

5.2.2

Settings for the control interface

If you wish to change the default settings for the control inputs and out-

puts (see section 5.2.3), proceed as follows:

The electronic circuitry of the 709 IC Pump is sensitive to electrostatic

charges. You should thus take the appropriate measures (earthing,

e.g. by touching metal) before opening the instrument.

1 Disconnect cables

Disconnect mains cable from mains connection plug 31 of the

709 IC Pump, also unplug all other cables connected to the rear

of the instrument.

2 Remove housing cover

Unscrew the two fastening screws 25 at the bottom of the

instrument and the two fastening screws 26 at the rear of the

instrument, remove housing cover 27.

3 Change settings

Change over blue jumpers at positions P12, P27, P28, P29 and

P30 of the main board (on left side of instrument when viewed

from back) using tweezers or flat-nosed pliers to match the

settings you wish (settings, see section 5.2.3):

P13 P12

P27 P28 P29 P30

1
2
3

4 Replace housing cover

Replace housing cover 27 and fasten with fastening screws 25

and 26.

5 Plug in cables

Connect mains cable to mains connection plug 31 and also plug

in all other cables.

Default settings