2 flow rate, 3 flow correction – Metrohm 709 IC Pump User Manual

Page 28

Advertising
background image

3 Operation

709 IC Pump

20

3.2 Flow

rate

The flow rate can be set in the range of 0.05

5.00 mL/min with the

pump drive switched on or off as follows:

1 Select "FLOW"

Press <SELECT> key 4 until the "FLOW" LED 10 lights up.

2 Set flow rate

Press <

> key 5 or <

> key 6 until the desired flow rate (in

mL/min) appears in display 1.

3 Store flow rate

Release <

> key 5 or <

> key 6: the value shown in display

1 at the time of release will be stored.

3.3 Flow

correction

Since the difference between displayed and actual flow rate can be

maximum 3 % owing to unavoidable tolerances in the manufacture of

mechanical parts, there is a possibility to adjust the 709 IC Pump after

installation of a new pump head. For this, a correction factor of 0.9

1.1

can be entered in the 709 IC Pump.

The correction factor is determined by measurement of the actual flow

rate with the aid of a measuring cylinder:

Correction factor =

Displayed flow rate (in mL/min)

Measured flow rate (in mL/min)

In the entry of the correction factor, proceed as follows:

1 Select flow correction

Press <SELECT> key 4 for ca. 3 s until all LEDs (10...13) go out

and the correction factor appears in display 1.

2 Set flow correction

Press <

> key 5 or <

> key 6 until the desired flow correc-

tion appears in display 1.

3 Store flow correction

Release <

> key 5 or <

> key 6: the value shown in display

1 at the time of release will be stored.

4 Return to standby mode

Press <SELECT> key 4 briefly, the pump is again in the

standby mode.