2 Installation

709 IC Pump

16

2.6 Mains

connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with a mains voltage set wrongly and/or

wrong mains fuse, there is a danger of fire!

2.6.1

Setting the mains voltage

Before switching on the 709 IC Pump for the first time, check whether

the mains voltage set on the instrument (can be read in the mains volt-

age selector 32) matches the local mains voltage. If this is not the case,

you must switch the mains voltage by pushing the mains voltage selec-

tor 32 with a screwdriver:



2.6.2 Fuses

The 709 IC Pump contains two fuses built in as standard, either of type

0.25 AT (for 230 V: 0.25 A, slow-blow, Metrohm ordering number

U.600.0010) or 0.5 AT (for 110 V: 0.5 A, slow-blow, Metrohm ordering

number U.600.0013). Ensure that the two fuses correspond to the fuse

type specified for the local mains voltage (see fuse data 28). To change

wrong or blown fuses, proceed as follows:

Ensure that the instrument is never put into operation with fuses of

another type, otherwise there is danger of fire!

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 31 of the

709 IC Pump.

2 Remove fuse cover

Using a screwdriver, lever out fuse cover 30 forwards until it

opens.

3 Check and change fuses if necessary

Carefully take the fuses installed for the desired mains voltage

out of the fuse holder and check its specifications:

100

120 V 0.5 A (slow-blow)

Metrohm No. U.600.0013

220

240 V 0.25 A (slow-blow) Metrohm No. U.600.0010

Change fuse if necessary and reinsert in fuse holder.

Position of the mains
voltage selector 32:

230V: 220

240 V

±

10%

115V: 110

120 V

±

10%

230V

115V