2 Installation

709 IC Pump

12

2.3.6

Filter unit Manufit

The 6.2821.000 Filter unit Manufit available as an option (see sec-

tion 6.3) serves to avoid contamination of the piston seals by abrasive

particles and can be used in the pressure range 0

50 MPa

(0

500 bar) together with steel capillaries. It is installed as follows (see

Fig. 5):

1 Prepare Manufit housing

Insert outlet capillary 55 with steel mesh holding end 53 into

Manufit housing 54.

Insert the 4 steel meshes 52 provided into the steel mesh

holding end 53.

Press the PTFE gasket 51 into the steel mesh holding end

53.

2 Prepare Manufit pressure screw

Insert inlet capillary 48 with counterpart end 50 into Manufit

pressure screw 49.

3 Assembly

Fit the two capillary end pieces 50 and 53 together.

Screw Manufit pressure screw 49 and Manufit housing 54

firmly together.

To replace contaminated steel meshes, proceed in the reverse order.

IC Separation
Center 733

IC Pump
709

48

49

50

51

52

53

54

55

Fig. 5: Filter unit Manufit (6.2821.000)

48 Inlet

capillary

49 Manufit pressure screw

50 Counterpart

end

51 PTFE gasket (6.2821.010)

52 4 Steel meshes (6.2821.020)

53 Steel mesh holding end

54 Manufit housing

55 Outlet capillary

Advertising