INSTALLATION, CONT.

Load Cell Cable Connection with RFI Suppression

CAUTION!

Disconnect any external load cell power supply before connecting

load cells to the indicator. Failure to do so will result in permanent damage to
the indicator.

Figure No. 2

AC Power

90-264 VAC

0.4 Amp

Scales 1,
2 and 3

I/O

(Serial,

Isolated

Inputs or

Outputs)

I/O

(Serial,

Isolated

Inputs or

Outputs)



To eliminate RFI, the load cell cable should be routed through the one of the special metallic
gland connector and the shield wire from the load cell cable must be connected to this gland
connector for grounding. Refer to Figure No. 2 for the gland connector layout.

1. Remove the 14 acorn nuts securing the rear

panel to the main housing.

2. Loosen and remove the metal gland connector

nut and remove the plastic insert.

3. Referring to Figure No. 3, route the single cable

from the load cell or load cell junction box
through the nut and plastic insert and into the
enclosure.

Figure No. 3



8200-M538-O1

y 225 Installation