CALIBRATION AND SETUP, CONT.

Dual-Point with Zero (First Zero) Calibration

on the

U.

CAL 1= (FIRST CALIBRATION WEIGHT)

With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and

CAL 1=0.0. This is the first of two calibration weights. It could be

ZERO (no load) or the TEST load (weight).

1. If the first calibration weight is to be ZERO (no load), press the ENTER key. The display

will flash

CALIBRATING... for a few seconds and then change to show CAL 2=.

2. If the first calibration weight is to be the TEST load (weight), place the weights on the

scale.

3. Using the numeric keys, enter the value of the TEST load (weight) and then press the

ENTER key.

4. The display will flash

CALIBRATING... for a few seconds and then change to

show

CAL 2=.


CAL 2= (SECOND CALIBRATION WEIGHT)

The display will show

CAL 2=XXXXXXX. This is the second of two calibration weights.

It could be ZERO (no load) or the TEST load (weight).

1. If the second calibration weight is to be ZERO (no load), press the 0 key and then the

ENTER key. The display will flash

CALIBRATING... for a few seconds and then

change to show

CAL 3=.

2. If the second calibration weight is to be the TEST load (weight), place the weights

scale.

3. Using the numeric keys, enter the value of the test load (weight) and then press the

ENTER key.

4. The display will flash

CALIBRATING... for a few seconds and then change to

show

CAL 3=.


CAL 3= (LAST CALIBRATION WEIGHT)

The display will show

CAL 3=. This weight is not used. Press the ENTER key to skip

CAL 3=. The display will change to show the SCALE 1 CALIBRATION MEN