CALIBRATION AND SETUP, CONT.

Your 225 indicator has been thoroughly tested and calibrated before being shipped to you. If
you received the indicator attached to a scale, calibration is not necessary. If the indicator is
being connected to a scale for the first time or recalibration is necessary for other reasons,
proceed as indicated.

Calibration and Setup of the indicator is accomplished entirely by the keypad and does not
require removing a calibration jumper or pressing a calibration switch

1

.

During the calibration and setup process it is necessary to enter operational parameters and
data using the indicator keypad.

• Pressing

the

ENTER key

U

without

U

entering a new value will retain the current setting and

return the display to the menu display.

• To change a setting, enter or select a new value and then press the ENTER key. The

new setting will be saved and the display will return to the menu display.

• Note that pressing the ←⎯ /ESC key on YES or NO prompts will “backup” to the

previous selection. Also note that on prompts requiring a value to be entered, pressing
the

←⎯ /ESC key will clear the value.

Alpha-numeric
Keys

Soft Keys

Navigation
Keys

Figure No. 13

CAUTION! The membrane keypad is not to be operated with pointed objects
(pencils, pens, fingernails, etc). Damage to keypad resulting from this practice
will NOT be covered under warranty.

1

If your local metrology laws require the use of physical sealing, to enter calibration and setup,

the calibration seal jumper must be removed. You may gain access to this jumper by
removing the fourteen (14) acorn nuts securing the rear panel. Refer to Figure No. 12 for the
location of the jumper.

8200-M538-O1

y 225 Installation

Page 34