INSTALLATION, CONT.

Optically Isolated Remote Inputs

Included with the I/O are 7 programmable inputs that may be used to remotely (up to 100 feet)
initiate various functions within the indicator. These inputs are accessed via a terminal block
(P17) on the back of the PC board (see Figure No. 11). The 7 inputs are defined as follows:

REMOTE INPUTS TERMINAL – (P17)

U

PIN NO.

U

U

Function

U

U

PIN NO.

U

U

Function

U

1

+ SRC(12 to 24VDC)

6

- - - - (Not Used)

2 ZERO

7 STOP

3 TARE

8 START

4 G/N

(Gross/Net)

9

DUMP

5 PRINT

10

GND/SHIELD

Remember that the input must be connected to GND to initiate the function.


AC Input Relay Board(s)

The AC Input Relay Board(s) are mounted in an external junction box for use with the 225
Indicator. The RB4-ACIN (115 VAC) or RB4-ACINV (230 VAC) contain one board and
supports 4 inputs (jumper selectable). The RB8-ACIN (115 VAC) or RB8-ACINV (230 VAC)
contain two boards and supports 7 inputs that are jumper selectable. The relay board used in
the 115 VAC versions is Cardinal p/n 8200-C324-0A. The 230 VAC version uses relay board
Cardinal p/n 8200-C324-1A. Connect the devices as shown in Figure No. 8.

INPUT RELAY TYPE IAC-5

90 to 140 VAC @ 6mA maximum for each plug-in relay

INPUT RELAY TYPE IAC-5A

180 to 280 VAC @ 6mA maximum for each plug-in relay

OUTPUT

5VDC @ 12mA from the 225 main pc board assembly P9
12VDC @ 12mA maximum from external source

CONNECTION

Removable plug-in screw terminals for up to 14 AWG wire

8200-M538-O1