INSTALLATION, CONT.

Preset Weight Comparator/Checkweigher Logic Level Output

If you so choose, you may use the logic level outputs from your Model 225 indicator’s preset
weight comparators or checkweigher to control peripheral devices used to manage the flow of
material or signal when the weight is within preset limits. Note that these outputs are at logic
level and cannot drive external devices directly. Solid-state relays can be used to accept the
logic level output from the 225 and in turn, drive the external device.

1. If the rear panel of the indicator has been removed, proceed to step 2. Otherwise, remove

the 14 acorn nuts securing the rear panel to main housing

2. Loosen the gland connector for the cable. Refer to Figure No. 2 for the gland connector

layout.

3. Slip the cable through the gland connector and into the enclosure.

4. Referring to Figure No. 4, remove 2 inches (not 5 to 6 inches) of the outer insulation jacket

5. Next, remove 1/4 inch of insulation from each of the wires.

6. Connect each of the wires to the Remote Outputs terminal block (P15 or P16) referring to

Figure No. 11 for terminal block locations.

7. To terminate a wire, use a small flat blade screwdriver and press down on the release bar

for the terminal. Insert the wire into the terminal opening. Remove the screwdriver,
allowing the release bar to return to its original position, locking the wire in place. See
Figure No. 5.

8. Repeat procedure until all wires are in place.

AC Output Relay Board(s)

The AC Output Relay Boards are mounted in an external junction box for use with the 225
Indicator. The RB4-ACOUT contains one board and supports 4 outputs (jumper selectable).
The RB8-ACOUT contains two boards and supports 8 outputs. The relay board used in both is
(Cardinal p/n 8539-C062-0A). Connect the devices to be controlled as shown in Figure No. 9
and 10.

The individual relays can be configured to be on (closed) or off (open) at weights under the
preset weight then switch at the preset weight from on-to-off or off-to-on by setting the under
weight condition to on or off during setup and calibration or setup review.

OUTPUT (closed)

28-240VAC @ 3A maximum for each plug-in relay

CONTROL INPUT

5VDC @ 12mA from the 225 main pc board assembly P8

CONNECTION

Removable plug-in screw terminals for up to 14 AWG wire



NOTE! All relays are the normally-open type that will open when power to indicator is lost.

8200-M538-O1