2 serial command configuration, 3 front panel configuration, 4 multi-range and multi-interval scales – Rice Lake 920i Installation Manual V3.10 User Manual

Page 27: 5 total scale configuration, Serial command configuration, Front panel configuration, Multi-range and multi-interval scales, Total scale configuration, Section 3.1.3

Advertising
background image

Configuration

21

3.1.2

Serial Command Configuration

The serial command set can be used to configure the

920i

indicator using either a personal computer,

terminal, or remote keyboard. Like

iRev

, serial

command configuration sends commands to the

indicator serial port; unlike

iRev

, serial commands can

be sent using any external device capable of sending

ASCII characters over a serial connection.
Serial commands duplicate the functions available

using the indicator front panel and provide some

functions not otherwise available. Serial commands

can be used to simulate pressing front panel keys, to

configure the indicator, or to dump lists of parameter

settings. See Section 9.0 on page 86 for more

information about using the serial command set.

3.1.3

Front Panel Configuration

Use the CONFIG submenu under the SCALES menu

to configure A/D scales. For example, in an indicator

with a single-channel A/D card installed in Slot 1, the

Scale Configuration display will show the A/D listed

(

Slot 1 Channel 1

) under the

AVAILABLE A/D’s

column.

Use the

left

navigation key to select the A/D, then

press the center softkey,

Add

. The A/D is then moved

to the Associated A/D’s column. If no other A/D’s are

listed in the

AVAILABLE A/D’s

column, the center

softkey changes to

Done

, as shown in Figure 3-2.

Press

Done

to exit the Scale Configuration display.

See Section 10.4 on page 108 for information about

configuring serial scales.

Figure 3-2. Scale Configuration Display

3.1.4

Multi-Range and Multi-Interval Scales

The

920i

supports multi-range and multi-interval

scales of either two or three ranges or intervals.
Multi-range scales provide two or three ranges, each

extending from zero to the maximum capacity

specified for the range, that can specify different scale

intervals (graduations). The scale interval changes as

the applied weight increases but does not reset to

lower range intervals until the scale returns to zero.
Multi-interval scales divide the scale into two or three

partial weighing ranges, each with different scale

intervals. The scale interval changes with both

increasing and decreasing loads applied.
To configure a multi-range or multi-interval scale, use

the SPLIT parameter to select 2RNG or 3RNG (for

multi-range scales), or 2INTVL or 3INTVL (for

multi-interval scales). Selecting a SPLIT value other

than OFF allows specification of decimal point,

display divisions, and maximum capacity for each

range or interval.

3.1.5

Total Scale Configuration

The output of two or more A/D scales or

iQUBE

systems can be configured to function as a total scale.

Once configured and calibrated, the total scale can be

used as a source for other system functions, including

streaming, setpoints, print formatting, and analog

output.
To set up a total scale from the indicator front panel,

use the scale configuration display (see Figure 3-2) to

select the A/D scales or

iQUBE

systems to configure

as a total scale. (Use the

Change Type

softkey to show

available A/D scales or

iQUBE

systems; use the right

navigation key to select the total scale sources.) In

iRev

, assign the total scale to an unused position then

select source scales from the existing A/D scales or

iQUBE

systems.

The FORMAT configuration of the total scale (see

Figure 3-5 on page 27) should match that of the

source scales. However, the value specified for the

total scale GRADS parameter should be specified as

the sum of the GRADS values for the source scales.

For example: if SCALE 1 is set to GRADS=10000,

SCALE 2 to GRADS=5000, SCALE 3 (the total

scale) should be set to 15000 grads.

The total scale will show an overrange indication if

the maximum capacity of any source scale is

exceeded, and show dashes if any source scale reads a

negative value. Source scales will respond to Tare and

Zero operations performed on the total scale.

Advertising